Skip to content

Model 6500-HA

HighAccuracyRVSMCompliantATEReady
Target Application

Aircraft instrumentation calibration.

Description

The 6500-HA (High-Accuracy) Air Data Calibrator can be used to test and calibrate altimeters, airspeed / Mach indicators, climb indicators, flight data recorders, air data computers, and EPR indicators.

This rack-mountable calibrator meets all RVSM requirements and requires only a pressure source and a vacuum source. The use of profiles makes it possible for the operator to run through a test using only a single key on the front panel keypad.

It can also be controlled externally as part of an ATE system.

Technical Reference

High-accuracy calibration.
RVSM compliant.

STATIC PORT

Pressure Function
Range: 0.1 to 38 inHg
Resolution: 0.0001 inHg
Accuracy: ±0.001 inHg
Units: inHg, mbar
Pressure Rate
Range: 0 to 30 inHg/min
Resolution: 0.001 inHg/min
Accuracy: ±1% of rate
Units: pressure-units/min
Altitude Function
Range: -4000 ft to 100,000 ft
Resolution: 1 ft
Accuracy: ±1 ft @ 0 ft, ±3 ft @ 35,000 ft, ±6 ft @ 50,000 ft, ±25 ft @ 80,000 ft
Units: feet, meters
Climb Function
Range: 0 ft/min to ±50,000 ft/min
Resolution: 1 ft/min
Accuracy: ±1% of rate of climb
Units: ft/min, meters/sec

PITOT PORT

Pressure Function
Range: 0.1 to 100 inHg
Resolution: 0.001 inHg
Accuracy: ±0.002 inHg
Pt Units: inHg, mbar, EPR
Qc Units: inHg, mbar
Pressure Rate
Range: 0 to 30 inHg/min
Resolution: 0.001 inHg/min
Accuracy: ±1% of rate
Units: pressure-units/min (EPR in inHg/min)
Airspeed Function
Range: 0 to 1100 knots
Resolution: 0.1 knots
Accuracy: ±1.0 knots @ 20 knots, ±0.3 knots @ 50 knots, <0.2 knots above 100 knots, <0.1 knots above 300 knots, ≤0.01 knots @ 1000 knots
Units: knots, Mach, km/hr, mph
Airspeed Rate
Range: 0 to 800 knots/min
Resolution: 0.1 knots/min (0.01 km/hr/sec)
Accuracy: ±2% of rate
Units: knots/min, km/hr/min, mph/min
Mach Function
Range: 0.0 to 5.0 Mach
Resolution: 0.001 Mach
Accuracy: ±0.001 above 0.2 Mach
EPR Function
Range: 0 to 199
Resolution: 0.001
Accuracy: ±0.002

SYSTEM

Power Requirement
AC Input
90-260 VAC, 47-440 Hz, 100 VA
External Interfaces
Interfaces
RS232 Serial, IEEE-488, Encoder (optional)
Remote
Model 6505 Remote (optional)
Dimensions & Weights
Main Unit
19.0" × 20.0" × 7.0" / 20 lbs
Mounting
4U Rack-mountable
Environmental Specs
Operating Temp
0°C to 50°C (20°C to 26°C for accuracy)
Storage Temp
-25°C to 75°C
Humidity
5 to 95% non-condensing

Need calibration
That ships?